Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/4436
Title: An RPC-PET prototype with high spatial resolution
Authors: Blanco, A. 
Carolino, N. 
Correia, C. M. B. A. 
Marques, R. Ferreira 
Fonte, P. 
González-Díaz, D. 
Lindote, A. 
Lopes, M. I. 
Macedo, M. P. 
Policarpo, A. 
Keywords: Gaseous detectors; Photodetectors; Resistive plate chambers (RPC); Positron emission tomography (PET); Small animal imaging
Issue Date: 2004
Citation: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 533:1-2 (2004) 139-143
Abstract: A small positron emission tomography system, based on the timing RPC technology has been built and tested. This first prototype is aimed at validating the expectations, derived from simulations, of a very high spatial resolution, which could be of value for the imaging of small animals. The system is composed of two counting heads, able to measure the photon interaction point in two dimensions, the transaxial dimension and the Depth of Interaction. Each head is composed of 16 independent stacked RPCs made from copper and glass (anode) electrodes. Point-like 22Na sources were so far reconstructed, using the standard filtered back-projection algorithm, with a spatial accuracy of 0.6 mm FWHM, free of parallax error.
URI: https://hdl.handle.net/10316/4436
DOI: 10.1016/j.nima.2004.07.016
Rights: openAccess
Appears in Collections:FCTUC Física - Artigos em Revistas Internacionais

Files in This Item:
File Description SizeFormat
filebbbc78d5ef6342f7a8b76c789d0f4038.pdf284.01 kBAdobe PDFView/Open
Show full item record

SCOPUSTM   
Citations

25
checked on Apr 22, 2024

WEB OF SCIENCETM
Citations

22
checked on Apr 2, 2024

Page view(s)

310
checked on Apr 16, 2024

Download(s)

322
checked on Apr 16, 2024

Google ScholarTM

Check

Altmetric

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.