Utilize este identificador para referenciar este registo: https://hdl.handle.net/10316/4436
Título: An RPC-PET prototype with high spatial resolution
Autor: Blanco, A. 
Carolino, N. 
Correia, C. M. B. A. 
Marques, R. Ferreira 
Fonte, P. 
González-Díaz, D. 
Lindote, A. 
Lopes, M. I. 
Macedo, M. P. 
Policarpo, A. 
Palavras-chave: Gaseous detectors; Photodetectors; Resistive plate chambers (RPC); Positron emission tomography (PET); Small animal imaging
Data: 2004
Citação: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 533:1-2 (2004) 139-143
Resumo: A small positron emission tomography system, based on the timing RPC technology has been built and tested. This first prototype is aimed at validating the expectations, derived from simulations, of a very high spatial resolution, which could be of value for the imaging of small animals. The system is composed of two counting heads, able to measure the photon interaction point in two dimensions, the transaxial dimension and the Depth of Interaction. Each head is composed of 16 independent stacked RPCs made from copper and glass (anode) electrodes. Point-like 22Na sources were so far reconstructed, using the standard filtered back-projection algorithm, with a spatial accuracy of 0.6 mm FWHM, free of parallax error.
URI: https://hdl.handle.net/10316/4436
DOI: 10.1016/j.nima.2004.07.016
Direitos: openAccess
Aparece nas coleções:FCTUC Física - Artigos em Revistas Internacionais

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato
filebbbc78d5ef6342f7a8b76c789d0f4038.pdf284.01 kBAdobe PDFVer/Abrir
Mostrar registo em formato completo

Citações SCOPUSTM   

25
Visto em 22/abr/2024

Citações WEB OF SCIENCETM

22
Visto em 2/abr/2024

Visualizações de página

311
Visto em 23/abr/2024

Downloads

323
Visto em 23/abr/2024

Google ScholarTM

Verificar

Altmetric

Altmetric


Todos os registos no repositório estão protegidos por leis de copyright, com todos os direitos reservados.