Please use this identifier to cite or link to this item:
https://hdl.handle.net/10316/4361
Title: | High-rate operation of the Micro-Hole and Strip Plate gas detector | Authors: | Veloso, J. F. C. A. Caldas, C. C. Oliveira, C. A. B. Azevedo, C. D. R. Santos, J. M. F. dos Breskin, A. Chechik, R. |
Keywords: | Count rate | Issue Date: | 2007 | Citation: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 580:1 (2007) 362-365 | Abstract: | We report the performance of the Micro-Hole-and-Strip-Plate (MHSP) operated in xenon, under high-flux irradiation of a 30-kV X-ray Cu-tube. Stable operation in pulse-counting mode for gas gains above 104 was measured up to counting rates of 5×105 Hz/mm2. A gain variation of less than 5% was observed for gains of 104, while the energy resolution for 8-keV X-rays degrades from ~16% to ~18%. No visible degradation was observed on the MHSP irradiated area. | URI: | https://hdl.handle.net/10316/4361 | DOI: | 10.1016/j.nima.2007.05.179 | Rights: | openAccess |
Appears in Collections: | FCTUC Física - Artigos em Revistas Internacionais |