Title: Mass Calibration of the Coulter Counter Model ZM
Authors: Figueiredo, M. Margarida 
Rasteiro, M. Graça 
Santos, Cristina 
Monteiro, Cristina 
Issue Date: 1991
Citation: Particle and Particle Systems Characterization. 8:1-4 (1991) 294-296
Abstract: The calibration of electrical sensing zone instruments is normally achieved by using spherical particles with a certified size. An alternative and more fundamental procedure, known as mass or self-calibration, is to use particles of the material under test.This work concerns the mass calibration technique, in particular the equations used to calculate the mass calibration constant. It is demonstrated that some of the published expressions are inconsistent. An expression particularly suitable for the Coulter Counter Model ZM has been derived and validated. Some experiments were also performed using irregular particles, in order to compare both calibration methods.
URI: https://hdl.handle.net/10316/8201
DOI: 10.1002/ppsc.19910080153
Rights: openAccess
Appears in Collections:FCTUC Eng.Química - Artigos em Revistas Internacionais