Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/5030
Title: Use of an acoustic wave sensor to follow lead absorption by porcine skin
Authors: Veríssimo, Marta I. S. 
Pais, Alberto A. C. C. 
Oliveira, João A. B. P. 
Gomes, M. Teresa S. R. 
Keywords: Bulk acoustic wave sensor; Stratum corneum
Issue Date: 2008
Citation: Sensors and Actuators B: Chemical. 128:2 (2008) 450-454
Abstract: A new acoustic wave sensor capable of detecting and quantifying lead is presented, and fully characterized in terms of sensitivity, selectivity, and linear calibration range. The sensor was used to quantify lead absorption by pig stratum corneum and to follow in real time, both the absorption and desorption of lead by the skin. Values of lead absorption of 29.6 mg Pb2+/g of stratum corneum were obtained. Lead in solution was also quantified by atomic absorption spectrometry (AAS). The results obtained by both methods, AAS and the sensor, were compared and no statistical differences in the accuracy and precision of the results were found.
URI: https://hdl.handle.net/10316/5030
DOI: 10.1016/j.snb.2007.07.011
Rights: openAccess
Appears in Collections:FCTUC Química - Artigos em Revistas Internacionais

Files in This Item:
File Description SizeFormat
file7bf2e1df62d94399a95aba690f8158fd.pdf302.62 kBAdobe PDFView/Open
Show full item record

SCOPUSTM   
Citations

3
checked on Apr 1, 2024

WEB OF SCIENCETM
Citations

3
checked on Apr 2, 2024

Page view(s)

349
checked on Apr 16, 2024

Download(s)

279
checked on Apr 16, 2024

Google ScholarTM

Check

Altmetric

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.