Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/5005
DC FieldValueLanguage
dc.contributor.authorFelix, Fabiana S.-
dc.contributor.authorBrett, Christopher M. A.-
dc.contributor.authorAngnes, Lúcio-
dc.date.accessioned2008-09-01T15:02:18Z-
dc.date.available2008-09-01T15:02:18Z-
dc.date.issued2008en_US
dc.identifier.citationTalanta. 76:1 (2008) 128-133en_US
dc.identifier.urihttps://hdl.handle.net/10316/5005-
dc.description.abstractFlow injection analysis (FIA) using a carbon film sensor for amperometric detection was explored for ambroxol analysis in pharmaceutical formulations. The specially designed flow cell designed in the lab generated sharp and reproducible current peaks, with a wide linear dynamic range from 5 × 10-7 to 3.5 × 10-4 mol L-1, in 0.1 mol L-1 sulfuric acid electrolyte, as well as high sensitivity, 0.110 A mol-1 L cm-2 at the optimized flow rate. A detection limit of 7.6 × 10-8 mol L-1 and a sampling frequency of 50 determinations per hour were achieved, employing injected volumes of 100 [mu]L and a flow rate of 2.0 mL min-1. The repeatability, expressed as R.S.D. for successive and alternated injections of 6.0 × 10-6 and 6.0 × 10-5 mol L-1 ambroxol solutions, was 3.0 and 1.5%, respectively, without any noticeable memory effect between injections. The proposed method was applied to the analysis of ambroxol in pharmaceutical samples and the results obtained were compared with UV spectrophotometric and acid-base titrimetric methods. Good agreement between the results utilizing the three methods and the labeled values was achieved, corroborating the good performance of the proposed electrochemical methodology for ambroxol analysis.en_US
dc.description.urihttp://www.sciencedirect.com/science/article/B6THP-4RW434G-1/1/d40a3f00a2eaccb84817b064f601d0f2en_US
dc.format.mimetypeaplication/PDFen
dc.language.isoengeng
dc.rightsopenAccesseng
dc.subjectFlow injection analysisen_US
dc.subjectCarbon film electrodesen_US
dc.subjectAmbroxolen_US
dc.subjectAmperometryen_US
dc.subjectPharmaceutical productsen_US
dc.titleFlow injection analysis using carbon film resistor electrodes for amperometric determination of ambroxolen_US
dc.typearticleen_US
dc.identifier.doi10.1016/j.talanta.2008.02.005-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.openairetypearticle-
item.cerifentitytypePublications-
item.grantfulltextopen-
item.fulltextCom Texto completo-
item.languageiso639-1en-
crisitem.author.orcid0000-0002-1972-4434-
Appears in Collections:FCTUC Química - Artigos em Revistas Internacionais
Files in This Item:
File Description SizeFormat
file6c1c4045afc4415a88e6be162c8fde11.pdf471.82 kBAdobe PDFView/Open
Show simple item record

SCOPUSTM   
Citations

18
checked on Apr 1, 2024

WEB OF SCIENCETM
Citations

17
checked on Apr 2, 2024

Page view(s)

344
checked on Apr 16, 2024

Download(s)

337
checked on Apr 16, 2024

Google ScholarTM

Check

Altmetric

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.