Resumo: A set of 35 photomultiplier tubes (ETL D730/9829Q), intended for use in the ZEPLIN III Dark Matter detector, was tested from room temperature down to -100°C, with the aim of confirming their suitability for detecting xenon scintillation light at 175 nm while immersed in the cryogenic liquid. A general improvement of both gain and quantum efficiency at the xenon scintillation wavelength was observed with cooling, the best combined effect being 40%, while little change was noted in the timing properties and dark current. Saturation of response due to accumulation of charge in the resistive bialkali photocathodes was seen at an average photocurrent of 108 photoelectrons/s for the device with best quantum efficiency, whereas an order of magnitude higher current was required to saturate the least sensitive one. Variations in photocathode thickness from tube to tube could account for this behaviour, as well as the fact that the quantum efficiency improves the most for devices with poorest efficiency at room temperature.
URI: https://hdl.handle.net/10316/4452
DOI: 10.1016/j.nima.2003.10.094
Direitos: openAccess
Aparece nas coleções:FCTUC Física - Artigos em Revistas Internacionais

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato
file105fec58d36a499da09afee71fbcae43.pdf334.26 kBAdobe PDFVer/Abrir
Mostrar registo em formato completo

Citações SCOPUSTM   

28
Visto em 1/abr/2024

Citações WEB OF SCIENCETM

21
Visto em 2/abr/2024

Visualizações de página 50

440
Visto em 16/abr/2024

Downloads

284
Visto em 16/abr/2024

Google ScholarTM

Verificar

Altmetric

Altmetric


Todos os registos no repositório estão protegidos por leis de copyright, com todos os direitos reservados.