Title: Activity measurements of 192Ir solid sources using a well-type ionization chamber
Authors: Gonçalves, J. A. C. 
Botelho, S. 
Pascholati, P. R. 
Ridenti, M. A. 
Fraga, M. M. F. R. 
Camara, J. R. 
Calvo, W. A. 
Bueno, C. C. 
Keywords: Ionization chambers; Activity; Radioactive decay
Issue Date: 2007
Citation: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 580:1 (2007) 222-225
Abstract: This work presents the preliminary activity measurements of 192Ir solid sources, applied in industrial gammagraphy, produced by the Sealed Sources Laboratory in the Radiation Technology Center (CTR) at IPEN-CNEN/SP. The pellets' activity measurements were carried out with a well-type ionization chamber developed at CTR. The chamber current-voltage response was studied as a function of 192Ir activities ranging from 27 GBq (0.75 Ci) to 4.1 TBq (110 Ci), under constant geometrical conditions. The results obtained have shown both the linearity of the ionization current and the absence of appreciable recombination or diffusion losses.
URI: https://hdl.handle.net/10316/4356
DOI: 10.1016/j.nima.2007.05.089
Rights: openAccess
Appears in Collections:FCTUC Física - Artigos em Revistas Internacionais

Files in This Item:
File Description SizeFormat
fileddab0c763cae4b1eb090384e3442fa52.pdf187.02 kBAdobe PDFView/Open
Show full item record

SCOPUSTM   
Citations

2
checked on Apr 1, 2024

WEB OF SCIENCETM
Citations

1
checked on Apr 2, 2024

Page view(s)

336
checked on Apr 16, 2024

Download(s) 50

349
checked on Apr 16, 2024

Google ScholarTM

Check

Altmetric

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.